ValleyLab Force FX [Refurbished]


Price:
Sale price$3,900.00

Description

The Valleylab Force FX is an isolated output electrosurgical generator that provides the power for cutting, desiccating, and fulgurating tissue during surgery. The generator is used for bipolar and monopolar electrosurgical procedures. You can use handset or footswitch controls to activate the generator. Receptacles are marked to ensure the correct connection of accessories. Indicators alert you when the generator is activated. It allows you to adjust the activation tone volume and reset the mode and power settings used previously. An RF activation port, RS-232 serial port, and expansion port are also included. It can be used in conjunction with the Valleylab Force GSU System and the Valleylab Electroshield Monitor.

Valleylab Force FX Specifications:

General

Dimensions

  • Height: 4.38” (11.1 cm)
  • Width: 14” (35.6 cm)
  • Depth: 17” (43.9 cm)
  • Weight: < 18 lbs. (< 8.1 kg)

Power

  • PER: 98
  • Input Power Requirements: Operating range is 85 to 132 AC volts. The maximum current is 7 amperes in Cut and 4 amperes in Coag.
  • High Frequency Leakage
    • Bipolar: Less than 60 mArms
    • Monopolar: Less than 150 mArms

Output Waveforms

  • Bipolar
    • Precise: 470 kHz sinusoid
    • Standard: 470 kHz sinusoid
    • Macro: 470 kHz sinusoid
  • Monopolar Cut
    • Low: 390 kHz sinusoid. Similar to the Pure Cut mode except the maximum voltage is limited to a lower value.
    • Pure: 390 kHz sinusoid
    • Blend 390 kHz bursts of sinusoid, recurring at 27 kHz intervals. 50% duty cycle envelope.
  • Monopolar Coag
    • Desiccate: 240 kHz sinusoid repeated at 39 kHz. 8% duty cycle.
    • Fulgurate: 390 kHz damped sinusoidal bursts with a repetition frequency of 30 or 57 kHz into 500 ohms
    • Spray: 390 kHz damped sinusoidal bursts with a randomized repetition centered at 28 kHz. Frequencies include 21kHz < f < 35 kHz. Output is further modulated by a random 250 Hz envelope with a variable duty cycle. Output power changes by less than 15% or 5 watts, whichever is greater, as the line voltage varies from 104-132 volts and 208-264 volts (at rated load)

Operating Parameters

  • Ambient Temperature Range: 10° to 40° C (50° to 104° F)
  • Relative Humidity: 30 % to 75%, noncondensing
  • Atmospheric pressure: 700 to 1060 millibars
  • Warm -up time: If transported or stored at temperatures outside the operating temperature range, allow one hour for the generator to reach room temperature before use.

Transport and Storage

  • Ambient Temperature range: -40° to 70° C (-40° to 158°F)
  • Relative humidity: 25% to 85% noncondensing
  • Atmospheric Pressure: 500 to 1060 minibars

REM Contact Quality Monitoring System

  • Measurement Frequency: 80 kHz ± 10 kHz
  • Measurement Current: Less than 10 μA
  • Acceptable Resistance Ranges: REM pad — 5-135 ohms; Non-REM pad — less than 20 ohms
  • Acceptance range is 5-135 ohms after REM PolyHesive II return electrode is applied.
  • Adaptive REM: REM trip is baseline impedance plus 40%. For example, if the baseline impedance is 30 ohms, the upper level trip is approximately 42 ohms. If the pad-patient impedance falls below the baseline impedance, a new baseline is established.

Features:

  • Instant Response technology that ensures the power delivered remains virtually constant, regardless of the tissue type.
  • Three internal microcontrollers that reduce system reaction time and increase the system’s processing speed.
  • Capacitive coupling is reduced by 30-50 percent when using Instant Response TM technology. This reduction is achieved by limiting the RMS voltage and the high-frequency harmonics. Lower voltage means less neuromuscular stimulation and more precise delivery of energy to reduce collateral damage.
  • Spray coagulation voltage of no more than 9000 volts for broad, but superficial coagulation with limited capacitive coupling.
  • Compatible with the exclusive Computer Motion Hermes Voice Command System. 