Description
The Puritan Bennett 980 is an advanced, full-featured respiratory ventilator designed to help patients breathe safely and comfortably. Using the ventilator’s unique capabilities, clinicians can synchronize the Puritan Bennett 980 with patient needs for a comprehensive, efficient caregiving strategy.
Specifications:
- Brand: Nellcor Puritan Bennett
- Model: Puritan Bennett 980
- Category: Ventilator
- Tidal Volume: 2 to 315 mL
- Ventilation Mode: (A/C), (SIMV), SPONT), BiLevel, (CPAP)
- Respiratory Rate: 1.0 1/min to 150 1/mi
- Rise Time: 1% to 100%
- Flow Pattern: Square or descending ramp
- Altitude: Operation: -411.5 m to 3048 m (-1,350 ft to 10,000 ft), Storage: 6,096 m max (20,000 ft max)
- Pendant Base Dimensions: 12.5 x 11.5 x 43.5 inch (32 x 30 x 109.22 cm)
- Pendant Base Weight: 76 lbs (34.5 kg) including BDU, GUI and primary battery; 60 lbs (27.2 kg) for BDU only
- Operating Temperature: 10°C to 40°C (50°F to 104°F)
- Display Size: 15 inch (38.1 cm)
- Display Type: Rotates 170° and tilts up to 45° from vertical
- Weight: 113 lbs (51.26 kg) including BDU and standard base; 69 lbs (31.3 kg) for GUI and primary battery BDU only
- Apnea Ventilation Mandatory Type: PC, VC, Neonatal default: PC
- Apnea Flow Pattern: Square or descending ramp
- Apnea Peak Inspiratory Flow (VMAX): 1.0 L/min to 30 L/min
- Apnea Tidal Volume: 3 mL to 315 mL
- Apnea Inspiratory Pressure: 5 cmH2O to 90–PEEP cmH2O
- Apnea Interval: 10 to 60 seconds or Off in CPAP, Neonatal default: 10 seconds
- Apnea Respiratory Rate: 2.0 1/min to 40 1/min, Neonatal default: 20 1/min
- Apnea O2%: 21% to 100% O2, Neonatal default: 40%
- Apnea I:E Ratio: ≤1.00:1
- Apnea Expiratory Time: 0.20 to 59.8 sec
- Disconnect Sensitivity: 1 L/min to 15 L/min, Neonatal default: 2 L/min for INV, 5 L/min in NIV
- Humidification Type: Heat-moisture exchanged (HME),Non-heated expiratory tube, heated expiratory tube
- Humidifier Volume: 100 mL to 1,000 mL
- Proximal Exhaled Tidal Volume: 0 mL to 500 mL
- Proximal Inspired Tidal Volume: 0 mL to 500 mL
- Inspiratory leak volume & Percentage: 0 mL to 9,000 mL, 0% to 100%
- Dynamic Resistance: 0 cmH2O/L/s to 100 cm22O/L/s
- Dynamic Compliance: 0 mL/cmH2O to 200 mL/cmH2O
- Inspiratory compliance: 0 to 1.00
- Apnea interval: 10 to 60 seconds or Off in CPAP, Neonatal default: 10 seconds
- Pressure Support: 0 to 70 cm H2O
- PEEP/interm. PEEP: 0 to 45 cmH2O
- Inspiratory Pressure Limit: 5 cmH2O to 90 cmH2O
- FiO2%: 21% to 100%
- Alarms: Exhaled tidal volume Low/high, Exhaled minute volume Low/high, inspired tidal volume High, Respiratory rate High, Circuit pressure Low/high
- Dimensions: 12.5 x 11.5 x 58 inch (32 x 30 x 148 cm ), including GUI screen
- Relative Humidity: 10% to 95% non-condensing
- Storage Temperature/ Humidity: 20°C to 70°C (-68°F to 158°F), 10% to 95% non-condensing
- Atmospheric Pressure: Operation: 70 kPa to 106 kPa (10.15 psi to 15.37 psi), Storage: 50 kPa to 106 kPa (7.25 psi to 15.37 psi)
- Sound Pressure Level: At a distance of 1 m, does not exceed 48 dBA at 5 L/min
- Plateau Time: 0.0 to 2.0 seconds
- Sensitivity: Expiratory sensitivity: 1% to 80%, Flow sensitivity: 0.1 L/min to 10 L/min, Neonatal default: 0.5 L/min
- I:E Ratio: 1:299 to 149:1
- Minimum expiratory time: ≥ 0.20 sec
Features:
- Versatile respiratory ventilator offers non-invasive SIMV and CPAP.
- Simple user interface with intuitive screen navigation capabilities.
- Customizable display for maximized efficiency.
- Ventilator assurance feature supports continued ventilation in the event of certain system failures.
- Integrated expiratory filtration system protects against cross contamination for both clinicians and patients.
- Clinicians can use the unit’s advanced tools to synchronize the ventilator with patient needs, offering the appropriate breathing support.
- Software enables spontaneous breathing and supports biphasic or airway pressure release ventilation.
- Proximal flow sensor to measure lower flows, pressures and tidal volumes in neonatal applications.
- Tube compensation accurately overcomes the work of breathing imposed by the artificial airway.
- Standby mode pauses ventilation and maintains settings while the patient is disconnected and auto-detects reconnection.
- Color touchscreen has large font option for enhanced visibility.
- Multiple USB ports and an HDMI output.
- Networking capabilities allow interfacing with existing EMR systems.
** All used goods are reconditioned by Tomlin Medical Pty Ltd. Used goods are not endorsed or supplied by their original manufacturer and are not subject to any manufacturer warranty. All used goods are subject to the limited warranty provided by Tomlin Medical Pty Ltd.**
Payment & Security
Your payment information is processed securely. We do not store credit card details nor have access to your credit card information.