Drager Oxylog 3000 [Refurbished]


Price:
Sale price$6,000.00

Description

Offering high ventilation performance with features such as AutoFlow integrated capnography and non-invasive Ventilation, the compact and robust Oxylog 3000 helps you transport your patients safely and provides feedback on the correctness of intubation and ventilation effectiveness. The Oxylog 3000 plus gives you the confidence to master even the most demanding situations.

Drager Oxylog 3000 Specifications:

General
  • Brand: Drager
  • Model: Oxylog 3000
  • Cagtegory: Ventilator
  • Patient Age Group: Adult
  • Tidal Volume: 0.05 to 2.0 L, BTPS
  • Ventilation Modes: VC-CMV, VC-AC, VC-SIMV, SpnCPAP, PC-BIPAP
  • Respiratory Rate: 2 to 60/min (VC-SIMV, PC-BIPAP), 5 to 60/min (VC-CMV, VC-AC), 12 to 60/min for apnoea ventilation

Settings

  • Oxygen concentration: 40 to 100 Vol.%
  • Inspiratory Pressure: PEEP +3 to +55 mbar
  • PEEP/CPAP: 0 to 20 mbar
  • Inspiratory time: 0.2 to 10 seconds
  • Inspiratory flow: 100 L/min @ supply pressures > 350 kPa / 51 PSI; 80 L/min @ supply pressures < 350 kPa / 51 PSI; 39 L/min @ supply pressures <270 kPa / 39 PSI
Monitoring
  • Peak Pressure: 0 to 35 mbar (relative to PEEP), slope adjustable in slow, standard, fast
  • Leakage: VTe. is approx. 60% lower than VTi (not applicable in NIV)
  • Incorrect patient hose: The ventilator detects if an incorrect patient hose type is connected

Ambient Conditions

  • Temperature (device) During Operation: -20 to +50°C for basic device, +10 to +40 °C for CO₂ sensor
  • Atmospheric pressure During Operation: 570 to 1200 hPa
  • Relative humidity During Operation: 5 to 95 % (no condensation)
  • Electromagnetic compatibility: Following IEC/EN 60601-1-2:2007, EN 794-3 and ISO 10651-3
  • Airworthiness: In accordance with RTCA DO-160F, sections 7, 8, 16.6, 18.3.1, 17, 19.31, 20, 21, 25
  • Mechanical strength: By MIL STD 810F, method 514.5

Performance

  • Display unit size: Pixels: 240 x 128, Visible area: 108 x 56 mm
  • Display Technology: Electro-luminescence (EL)
  • Curve display: Airway pressure Paw curve, flow curve, CO₂ curve (optional)

Operating Data

  • Current consumption: 24 V ±6 VDC
  • Dimensions: 290 x 184 x 175 mm
  • Weight: Approximately 5.8 kg (including internal battery)
  • Options: Integrated mainstream CO₂ measurement, Real-time data export via RS232, MEDIBUS protocol, AutoFlow: volume targeted - pressure controlled ventilation
  • Ventilation time ratio: 1:100 to 50:1
  • Trigger sensitivity: 1 to 15 L/min
  • Measured value display: MVe, FiO₂, RR, VTe, PEEP, Pmean, PIP, Pplat, MVesp, RRspon, etCO₂
  • CPR behavior: Pressure-limited, non-constant-volume ventilation during inspiration time when Pmax is reached
  • Patient hose types: Reusable adult hose (1,5 m / 3 m), Disposable adult hose (1,5 m / 3 m), Disposable pediatric hose (1,9 m)

Power Supply

  • Input voltage: 24 V ±6 VDC
  • AC/DC power pack: Input: 100 to 240 V~ / 50 to 60 Hz / 0.9 to 0.4 A~
  • DC/DC converter: Input: 12 / 24 / 28 VDC; 5 A / 2.5 A / 2.1 A
  • Operating time: Approximately 9.5 hours
  • Operation time: Approximately 7.5 hours
  • Battery Type: Lithium-ion battery
  • Charging time: Approximately 4 hours

Alarm

  • Airway Pressure: Paw high: Adjustable from 20 to 60, Paw low: When the pressure difference between Insp. and Exp. < 5 mbar or when the set pressure level is not reached
  • Apnea alarm time: When respiratory activity is no longer detected, adjustable time from 15 to 60 seconds
  • End-expiratory CO concentration (etCO₂): When the alarm limits for end-expiratory CO₂ concentration have been exceeded
  • Expiratory minute volume (MVe): When the alarm limits for expiratory minute volume have been exceeded

Additional Details

  • Gas supply: Medical Oxygen from a pipeline system or from an O₂ cylinder
  • Supply pressure: 270 kPa to 600 kPa at 100 L/min
  • Gas consumption for internal control: Average 0.5 L/min
  • Classification: Class IIb according to MDD 93/42/EEC
  • UMDNS-Code: 18-098
  • Additional settings for ventilation: Pressure support: in the ventilation modes VCSIMV, PC-BIPAP and SpnCPAP, Apnoea ventilation: in the ventilation mode SpnCPAP, AutoFlow (optional): in the ventilation modes VC-CMV, VC-AC and VC-SIMV, NIV: in the ventilation modes: SpnCPAP (/PS), PC-BIPAP (/PS), VC-CMV /AF, VC-AC /AF and VC-SIMV/AF
  • Special procedures: Inspiration hold, O₂ inhalation (optional), with an inhalation mask, 100% O₂

Features:

  • High-performance ventilation capability
  • Wide range of ventilation modes and autoflow
  • Advanced data export functionality
  • Perfectly portable & proven durability

Payment & Security

American Express Apple Pay Google Pay Mastercard PayPal Shop Pay Union Pay Visa

Your payment information is processed securely. We do not store credit card details nor have access to your credit card information.